Purity 99.37 % and endotoxin 7.53 EU/mg are unrelated — purity measures synthesis quality, endotoxin measures fill-step water quality. The combination here is coherent and expected together.
pegmgf_disperse
Polydisperse conjugates and the meaninglessness of a single purity number.
Section 1
Contribution
Comments technically on records submitted by others. Annotations are signed and are not moderated for agreement.
Records submitted
0
n=0
Annotations
2
n=2
Median purity submitted
suppressed
n=0
n<3 — see individual records
Suppliers covered
0
n=0
This handle has submitted no records. Annotators comment on other people's submissions and are not required to submit.
Section 2
Annotations
Signed technical comments on records. The index does not moderate annotations for agreement with its own arithmetic, and several of these disagree with it.
Endotoxin determined but the LAL method is not stated. Kinetic chromogenic and gel-clot give different results and different sensitivity. The value cannot be interpreted without knowing which.