AssayIndex
Reference

Glossary

238 terms across 10 categories — chromatography, mass spectrometry, content and mass, impurities and degradation, microbiology, peptide chemistry, statistics, quality systems, and this index's own conventions. Every technical term used on a record page resolves here.

238 terms10 categories
43 terms

Chromatography

Area percent
Purity expressed as the main peak's integrated area divided by the total integrated area of all peaks, times 100. A relative quantity. It is not a concentration and not a content.
Asymmetry factor (As)
Peak shape measure taken at 10 % of peak height: the width of the rear half divided by the width of the front half. Symmetric = 1.00. Above 1.50 the index marks integration unreliable.
see also Rule AX-R12
Backpressure
The pressure the pump develops to push mobile phase through the column. Rising backpressure across a sequence indicates a blocking frit or a fouled column inlet.
Baseline
The detector signal in the absence of an eluting analyte. On a gradient it drifts, and where the baseline is drawn determines the integrated area.
Capacity factor (k')
Retention relative to the unretained time: (t − t₀)/t₀. A k' below 1 means the compound is barely retained and is not usefully separated.
see also Void volume
Column dead time (t₀)
The time taken for an unretained species to traverse the column. Everything eluting at t₀ is unseparated.
see also Void volume
Core–shell particle
A stationary-phase particle with a solid core and a porous shell. Shorter diffusion path, sharper peaks, lower backpressure than a fully porous particle of the same size.
Drift
Slow monotonic change in the baseline across a run. Caused by mobile-phase absorbance changing with organic fraction, or by column temperature not being controlled.
see also Baseline
Drop-line integration
Separating two overlapping peaks with a vertical line dropped from the valley between them. Unbiased when the peaks are of comparable size.
Gradient
Programmed change in mobile-phase composition during a run. Gradient duration determines resolving power and is therefore part of the result, not metadata.
Gradient slope
Percentage points of organic solvent per minute. Lower slope, better resolution, longer run.
see also Gradient
Guard column
A short disposable column before the analytical column, protecting it from particulates and strongly retained matrix. Shifts retention slightly and must be accounted for.
HILIC
Hydrophilic-interaction liquid chromatography. Polar stationary phase, high-organic mobile phase. Retains polar analytes that reverse phase cannot hold.
Ion-pair chromatography
Adding an amphiphilic counter-ion to the mobile phase so a charged analyte forms a neutral pair with enough hydrophobic character to be retained on C18.
see also HILIC
Isocratic
Constant mobile-phase composition throughout the run. Simpler, and unsuitable for samples spanning a wide hydrophobicity range.
see also Gradient
Mobile phase
The liquid carrying the sample through the column. In peptide work typically water/acetonitrile with 0.1 % trifluoroacetic acid.
Peak apex
The point of maximum detector response within a peak. Where the UV spectrum and mass spectrum are usually taken.
Peak integration
Determining a peak's area by defining its start, end and baseline. The step where a purity figure is actually decided, and the one least often documented.
Peak purity
Evidence from a photodiode-array or mass detector that the spectrum is constant across a peak, indicating no co-eluting species with a different chromophore. Silent on diastereomers.
see also Diastereomer
Photodiode array (PDA)
A UV detector acquiring a full spectrum at every time point rather than one or two wavelengths. Enables peak-purity assessment and the 280/214 ratio.
see also Peak purity
Plate count (N)
Column efficiency: N = 5.54 (t/w½)². Higher is sharper. The index requires N ≥ 20,000 for a main peak.
Plate height
Column length divided by plate count. A physical measure of the efficiency per unit length, independent of column length.
Re-equilibration
Returning the column to starting conditions after a gradient. At least ten column volumes; short re-equilibration causes retention drift across a sequence.
Reporting threshold
The area percent below which peaks are not reported. The index uses 0.05 %. Excluded peaks are excluded from the total, so purity is systematically overstated by their sum.
Identification threshold
The level above which an impurity must be structurally identified rather than merely quantified. ICH Q3A sets it by daily dose.
Qualification threshold
The level above which an impurity must be shown to be biologically qualified. Not applicable to research material and included here for completeness.
Resolution (R)
Separation between two peaks: R = 2(t₂ − t₁)/(w₁ + w₂). R = 1.5 is baseline resolution. The index requires R ≥ 1.20 to the nearest named impurity.
see also Rule AX-R12 · Valley
Retention time
Time from injection to peak apex. A property of the compound, the column and the gradient — comparable across laboratories only after gradient normalisation.
Reverse-phase chromatography
Separation on a non-polar stationary phase, typically C18-bonded silica, with a polar mobile phase. Retains by hydrophobic interaction.
see also HILIC · C18
C18
Octadecyl-bonded silica: the standard reverse-phase stationary phase for peptides. Retention scales with the analyte's exposed hydrophobic surface, not with its mass.
Shoulder
An unresolved peak appearing as an asymmetric bulge on the leading or trailing edge of a larger peak. The main source of integration ambiguity.
Solvent front
The disturbance at the start of a chromatogram caused by the injection solvent. Compounds eluting within it cannot be integrated.
System suitability
Injections demonstrating the chromatographic system was performing acceptably when the sample was run. Required by every pharmacopoeia, documented on almost no research-peptide certificate.
Tailing factor (T)
Peak shape measure at 5 % of peak height: (a + b)/2a. USP convention; related to but not identical with the asymmetry factor.
Tangent skim
Integrating a small peak on the trailing edge of a much larger one by drawing a tangential baseline. The convention above a height ratio of about 10:1.
Trifluoroacetic acid (TFA)
Mobile-phase additive at 0.1 %. Gives excellent peptide peak shape and suppresses electrospray ionisation. Also the most common peptide counter-ion.
Valley
The minimum of the detector signal between two adjacent peaks. Valley height relative to the smaller peak determines whether the two can be reliably integrated separately.
Valley-to-valley integration
Baselining a small peak from valley to valley rather than to the true baseline. Assigns the overlap to the larger peak and understates the smaller one.
Void volume
The volume of mobile phase within the column. Determines the dead time. A void marker injection calibrates the retention axis and is cheap and rarely done.
Wavelength, detection
The UV wavelength at which absorbance is measured. 214 nm sees the amide backbone; 280 nm sees only aromatic residues. Purity figures at the two are not comparable.
Chromophore
The part of a molecule responsible for UV absorbance. In peptides: the amide bond below 230 nm, and Trp/Tyr near 280 nm.
Ghost peak
A peak arising from material retained on the column from a previous injection and released by a later gradient. Detected by a blank injection.
see also Carry-over
Carry-over
Sample residue transferred from one injection to the next, usually in the autosampler needle or loop. Quantified by a blank after a high-concentration injection.
see also Ghost peak
25 terms

Content & mass

Amino-acid analysis (AAA)
Acid hydrolysis followed by chromatographic quantification of the liberated residues. The reference method for peptide content where no standard of the intact peptide exists.
Anhydrous basis
A content figure recalculated as if the sample contained no water. Must be stated, or a mass balance will not close.
Assay
Quantification of an analyte against a material of known content. Distinct from area percent, which compares peaks within one chromatogram.
Counter-ion
The anion paired with a peptide's basic residues in the isolated salt. Usually trifluoroacetate or acetate. Mass in the vial that is not peptide.
Fill weight
The gross mass of material in a vial as weighed. Not the peptide mass — it includes water, counter-ion and residual solvent.
Gravimetric fill check
Weighing a vial before and after removal of contents to determine the actual fill. The only direct measurement of label-claim compliance.
Karl Fischer titration
Stoichiometric determination of water by reaction with iodine and sulfur dioxide. The reference method for residual moisture in lyophilised material.
Coulometric titration
A titration in which the titrant is generated electrochemically and quantified from the charge passed. For water: 10.71 coulombs per milligram.
Label claim
The mass of compound stated on the vial. Whether it refers to gross fill or to peptide content is almost never specified.
Mass balance
Water + counter-ion + residual + total peptide content = 100 % of cake mass. Exhaustive and mutually exclusive. The index closes it on every record where the terms exist.
Peptide content
The fraction of total mass that is peptide, %w/w. Distinct from purity, which is a fraction of detected chromatographic area.
Target-compound content
Total peptide content multiplied by chromatographic purity. The fraction of the vial that is the named compound. The number that determines recovered mass.
Recovered mass
Gross fill × target-compound content / 100. The milligrams of named compound actually obtainable from the vial.
Relative response factor
The detector response of an impurity per unit mass, relative to the parent. Assumed to be 1.00 in almost all purity determinations, and usually is not.
Residual moisture
Water remaining in a lyophilised cake. 1–4 % in well-dried material; 8–13 % for hygroscopic short peptides.
Residual solvent
Process solvent remaining in the product. DMF, DMSO, acetonitrile and dichloromethane are the usual candidates in peptide synthesis. Controlled by ICH Q3C.
GC headspace
Gas chromatography of the vapour above a heated sealed sample. The standard technique for residual solvents.
Salt-free basis
A content figure recalculated as if the sample contained no counter-ion. Like the anhydrous basis, must be stated explicitly.
Salt form
Which counter-ion the compound was isolated as. Trifluoroacetate is 114 Da per site; acetate is 60. At equal purity, acetate vials contain more peptide.
see also Counter-ion
Sulfated ash
The inorganic residue remaining after ignition with sulfuric acid. A measure of inorganic contamination and part of the mass balance.
see also Mass balance
Quantitative NMR (qNMR)
Absolute quantification by NMR against an internal standard of certified purity. Method-independent and rarely used in this market.
Nitrogen determination
Content estimation from total nitrogen, by Kjeldahl or combustion. An orthogonal check on AAA. Confounded by any nitrogen-containing excipient.
Ion chromatography
Separation of ions on an ion-exchange column with suppressed conductivity detection. The reference technique for counter-ion quantification.
see also Counter-ion
ICP-MS
Inductively coupled plasma mass spectrometry. Elemental analysis. Required for metal-complexed peptides and for elemental impurity profiles.
Elemental impurity
Trace metal contamination, controlled by ICH Q3D. Palladium, copper, nickel and chromium are the relevant risks in peptide synthesis. Not reported anywhere in this index.
see also ICP-MS
31 terms

Impurities & degradation

Acetylation
Addition of an acetyl group, +42.011 Da. Either a deliberate synthesis capping step or an unwanted side reaction.
see also Capping
Aggregate
A non-covalent assembly of two or more molecules. Dissociated by the acidic organic mobile phase of reverse-phase HPLC, so invisible to it. Seen by size-exclusion chromatography.
see also Fibrillation
Aspartimide
A cyclic imide formed at an Asp-X motif, −18.011 Da. Hydrolyses to a mixture of α- and β-aspartyl peptides, the latter mass-identical to the parent.
Asp-Gly motif
An aspartate followed by a glycine. The most aspartimide-prone sequence in peptide chemistry. Epitalon's defining analytical problem.
see also Aspartimide
β-Aspartyl isomer
A peptide in which the backbone connects through the aspartate side-chain carboxyl. Mass-identical to the parent and barely resolved. Requires a dedicated method.
see also Aspartimide
Capping
Deliberately acetylating unreacted amines during synthesis so failed couplings terminate as short capped species rather than continuing as deletion sequences.
Deamidation
Conversion of asparagine to aspartate or glutamine to glutamate, +0.984 Da. A storage degradation route, accelerated above pH 7.
see also Isoaspartate
Isoaspartate
The rearranged product of deamidation via a succinimide intermediate. Mass-identical to aspartate-containing peptide.
Deletion sequence
A peptide missing one or more residues from an incomplete coupling. Mass difference equals the missing residue mass. A synthesis finding: it does not form in a vial.
Diastereomer
A stereoisomer arising from inversion at one chiral centre. Same formula, same mass, same UV spectrum, often much less active. The impurity no mass spectrum finds.
Diketopiperazine
A cyclic dipeptide, −18.011 Da, formed by intramolecular attack at the N-terminus. Common in proline-rich short peptides.
see also Pyroglutamate
Dimer
Two molecules covalently linked, usually through a disulfide. Mass is roughly double the parent, minus 2.016 Da for a disulfide linkage.
Disulfide scrambling
Cysteines paired in an incorrect combination. Mass-identical to correct pairing and only detectable by peptide mapping or comparison against a folded reference.
Open-chain form
A cyclic peptide with its disulfide reduced, +2.016 Da. A handling artefact, resolved chromatographically, and functionally inactive.
Epimerisation
Inversion of a single stereocentre during synthesis, typically during carboxyl activation. Produces a diastereomer.
Racemisation
Loss of stereochemical purity toward a 50:50 mixture. In stepwise peptide synthesis the practical concern is epimerisation at individual residues.
see also Epimerisation
Fibrillation
Ordered self-assembly into amyloid-like fibrils. Glucagon is the fastest fibrillator in this index; native amylin is why pramlintide exists.
see also Aggregate
Fmoc
Fluorenylmethyloxycarbonyl, the standard temporary amine protecting group in solid-phase synthesis. Residual Fmoc appears at +222.068 Da.
Oxidation
Addition of oxygen. Methionine to sulfoxide (+15.995) then sulfone (+31.990); tryptophan to oxindole and beyond; cysteine to sulfonic acid.
Oxidation ladder
The series of peaks from sequential oxidation of multiple oxidisable residues: parent, mono-sulfoxide, bis-sulfoxide. Diagnostic — a single early peak with no second rung is probably something else.
see also Oxidation
Sulfoxide
The +15.995 Da methionine oxidation product. Elutes earlier than the parent on reverse phase. A handling finding, not a synthesis finding.
see also Oxidation
Pyroglutamate
Cyclisation of an N-terminal glutamine or glutamate, −17.027 or −18.011 Da. Also the native N-terminus of GnRH, where acid stress opens it to +18.011.
Regioisomer
An isomer differing in the position of a substituent, for example an acyl group on a different side chain. Mass-identical.
see also Diastereomer
Impurity fingerprint
The vector of named impurities and their levels in a lot. Characteristic of one manufacturing run; used by the index to identify shared bulks.
Shared bulk
One physical batch of material sold by two or more suppliers under different lot numbers. Normal in a contract-synthesis market; means two results are one measurement.
TFA adduct
A trifluoroacetylated peptide, +95.982 Da, formed during cleavage or from prolonged exposure to trifluoroacetic acid.
Truncation
A peptide missing residues from one terminus. Distinguished from an internal deletion by the mass difference matching a contiguous run of residues.
tBu residual
Incompletely removed tert-butyl side-chain protection, +56.063 Da. Indicates a short or weak cleavage step.
see also Fmoc
Forced degradation
Deliberately stressing a sample under acid, base, oxidative, thermal and photolytic conditions to prove the method can resolve degradants.
Stability-indicating method
A method demonstrated to separate the analyte from all its degradation products. A stability claim from a non-stability-indicating method means nothing.
30 terms

Index conventions

n=
The number of records a statistic was computed from. Displayed as a badge beside every statistic on this site; amber below 8, suppressed below 3.
Provenance
How a sample was obtained: buyer-sourced, supplier-supplied or laboratory-retained. A first-class field in this index because it changes the weight of the evidence.
Buyer-sourced
Purchased through the supplier's ordinary retail channel by the submitter, with no supplier involvement. The strongest provenance class here.
see also Provenance
Supplier-supplied
Provided by the supplier for testing. The supplier chose the vial. Reads a mean 0.31 pp higher than buyer-sourced across this index.
Assay report
The primary object in this index: one analysis of one sample of one batch by one laboratory. Everything else on the site is a rollup of these.
see also Batch rollup
Batch rollup
All reports in the index for one physical lot, with a reproducibility assessment across independent tests.
Within-batch spread
The range of independent purity results on one lot. A statement about bulk homogeneity and fill control, distinct from median purity.
see also Batch rollup
COA divergence
The supplier's declared purity minus the median of independent measurements on the same lot. Flagged above 1.5 pp by rule AX-R05.
see also Rule AX-R05
Documentation-completeness score
The index's 0–100 rubric for what a supplier's certificates actually state. Rubric published in full at /suppliers/methodology/.
Gradient normalisation
Converting a retention time measured on one gradient duration to the index reference scale, so retentions from different laboratories can be compared.
Index reference method RG-2
The index's nominated reverse-phase method: C18 2.6 µm, 0.1 % TFA, 5→65 % acetonitrile over 30 min at 40 °C, 214 nm. All retention windows are stated against it.
Method limited
A verification status: the method used does not meet the index specification for the compound class. The result is published with the limitation named.
Partial
A verification status: a measurement the compound requires is missing — a metal determination on a complex, an anomeric ratio on NAD+.
Purity only
A verification status: no identity confirmation was ordered. The record describes homogeneity and does not establish what the material is.
Verification status
The index's summary of how complete a record is: verified, identity consistent, purity only, method limited, partial, or not applicable.
Rule AX-R01
Mass accuracy outside the class limit. Records identity as not established rather than failed, because most instances are calibration problems.
see also Mass accuracy
Rule AX-R02
A certificate's declared purity and declared total impurities do not sum to 100 %. A documentation finding: the certificate is not interpretable as written.
Rule AX-R03
A certificate reports a largest single impurity exceeding its total impurities. Arithmetically impossible; a template error.
Rule AX-R04
Repeated round-number purity declarations from one supplier. Requires a pattern of at least three, because one round result is unremarkable.
Rule AX-R05
Declared purity more than 1.5 pp above the median of at least two independent tests on the same lot. The index's most consequential rule.
Rule AX-R06
Recovered mass outside ±10 % of label claim. The rule that fires most often and matters most to anyone buying by the milligram.
Rule AX-R07
Declared water, counter-ion, peptide content and residual do not account for 100 % of mass. Usually an unstated reporting basis.
see also Mass balance
Rule AX-R08
Endotoxin above the index screening threshold of 5.0 EU/mg. A screening figure, not a pharmacopoeial limit.
Rule AX-R09
Impurity fingerprints matching across suppliers, indicating one bulk under two labels. Not improper; means two results are one measurement.
Rule AX-R10
A supplier-supplied sample reading more than 0.8 pp above buyer-sourced material from the same lot.
Rule AX-R11
Retention time outside the compound window after gradient normalisation. Two thirds of instances are undocumented method differences.
Rule AX-R12
Peak asymmetry above 1.50 or resolution below 1.20 to the nearest impurity. Makes the reported purity a range rather than a number.
Submission
A report offered to the index by a buyer or laboratory. Becomes a record after normalisation and provenance classification.
see also Assay report
Annotation
A short signed technical comment attached to a record. The index's only community surface — there is no forum and no voting.
Suppressed statistic
A statistic the index declines to publish because n < 3. The individual values are shown instead.
see also n=
21 terms

Mass spectrometry

Adduct
An ion formed by association with a cation other than a proton — sodium at +21.982, potassium at +37.956. Not an impurity; the same molecule with a different cation.
Average mass
The abundance-weighted mean mass over the whole natural isotope distribution. What a low-resolution instrument reports. About 600 ppm above the monoisotopic mass for a peptide.
Base peak
The most intense peak in a spectrum. For a 4 kDa peptide by electrospray it is usually the 4+ charge state.
Charge state (z)
The number of protons carried by an ion. Observed m/z = (M + z × 1.007276)/z. Two adjacent charge states determine z without assumption.
Charge-state envelope
The distribution of intensities across charge states. Its shape carries conformational information: compact molecules ionise at lower charge.
Deconvolution
Recovering a neutral mass from a set of observed m/z values across charge states. Self-checking: two charge states give two independent estimates.
Electrospray ionisation (ESI)
Producing gas-phase ions from solution by applying a high potential to a fine spray. Multiply charges large molecules and is compatible with LC.
High-resolution mass spectrometry
Sufficient resolving power to distinguish species differing by a small fraction of a dalton. Required to exclude 1 and 2 Da alternatives.
Ion suppression
Reduction in analyte ionisation caused by a co-eluting species or a mobile-phase additive. Trifluoroacetic acid is a strong suppressor.
Isotope pattern
The set of peaks arising from natural isotopic variants of a molecule. Its spacing gives the charge state directly: 1/z daltons.
Lock mass
A reference compound infused continuously so calibration drift can be corrected in real time. Its absence limits mass accuracy to the last external calibration.
see also Mass accuracy
Mass accuracy
Difference between measured and theoretical mass, in ppm of the theoretical value. A relative measure — the same ppm is a different absolute error at different masses.
see also ppm · Rule AX-R01
Mass defect
The difference between a nominal integer mass and the exact mass. Constrains which elemental compositions are plausible for an observed impurity.
Monoisotopic mass
The mass of the isotopologue in which every atom is its most abundant isotope. All theoretical masses on this site are monoisotopic.
MS/MS
Tandem mass spectrometry: selecting an ion, fragmenting it, and measuring the fragments. Required to distinguish isobaric species such as pyroglutamate from aspartimide.
ppm
Parts per million. Used for mass accuracy: ppm = (observed − theoretical)/theoretical × 10⁶.
see also Mass accuracy
Resolving power
A mass spectrometer's ability to separate two close masses, expressed as m/Δm. A quadrupole gives ~1,000; a TOF 20,000–50,000.
Space charge
Mutual repulsion between ions at high density, shifting measured m/z. Why an overloaded injection can read a different mass from an on-scale one.
see also Mass accuracy
Time-of-flight (TOF)
A mass analyser measuring the time for ions to traverse a flight tube. High resolving power; calibration is temperature-sensitive.
Unit resolution
Resolving power sufficient to separate masses one dalton apart but not to determine exact mass. Cannot exclude a +1 or +2 Da alternative species.
Quadrupole
A mass analyser using oscillating electric fields. Robust, inexpensive, unit resolution.
13 terms

Microbiology

Bacterial endotoxin
Lipopolysaccharide from gram-negative bacterial outer membranes. Pyrogenic, filter-passing, autoclave-stable, and invisible to every chromatographic method.
Endotoxin unit (EU)
The activity unit for bacterial endotoxin, defined against a reference standard. Roughly 0.1–0.2 ng of E. coli LPS per EU.
LAL
Limulus amebocyte lysate. The horseshoe-crab-derived reagent whose clotting cascade detects endotoxin. Available as gel-clot, turbidimetric and chromogenic assays.
Gel-clot
A limit-test LAL format: a firm gel forms or it does not. Gives a pass or fail against a stated sensitivity, never a number.
see also LAL
Kinetic chromogenic
A quantitative LAL format reading the onset time of a colour change against a standard curve. Sensitive to 0.005 EU/mL.
see also LAL
Recombinant factor C (rFC)
A recombinant alternative to LAL using the cascade's first enzyme. Equivalent sensitivity, no horseshoe crabs.
see also LAL
Inhibition/enhancement control
A spike of standard endotoxin added to the sample to show the matrix does not interfere. Must recover at 50–200 % or the result is invalid.
see also LAL
Maximum valid dilution
The greatest dilution at which an endotoxin limit can still be detected. Sets how far a sample may be diluted to overcome interference.
Threshold pyrogenic dose (K)
5.0 EU per kilogram per hour for parenteral routes, 0.2 EU/kg for intrathecal. Converting it to a per-milligram limit requires the dose.
Sterility test
Incubation of the article or its filtrate in growth media for at least 14 days. Destructive, low-powered, and not a statement about the lot.
Membrane filtration
The preferred sterility technique: filter the article through 0.45 µm, then incubate the membrane in media.
Method suitability (sterility)
Demonstration that the article does not inhibit growth of challenge organisms. Omitting it on an antimicrobial peptide produces a false pass.
Bioburden
The viable microbial count in a non-sterile article. Distinct from sterility, which is a presence/absence test.
22 terms

Peptide chemistry

Aib
α-Aminoisobutyric acid. An achiral, sterically hindered residue used to resist enzymatic cleavage. Present in every incretin co-agonist in this index; its hindered couplings drive epimerisation nearby.
Acylation
Attachment of a fatty-acid chain to a side chain, conferring albumin binding and a long half-life. The defining modification of the long-acting GLP-1 analogues.
see also AEEA linker
AEEA linker
2-(2-(2-aminoethoxy)ethoxy)acetic acid. The spacer between the peptide backbone and the fatty-acid chain in semaglutide. Its loss gives the diagnostic −145 Da impurity.
see also Acylation
Coupling efficiency
The fraction of resin-bound chains that react at each synthesis step. At 99.5 % per step, a 40-mer is 82 % full-length before purification.
C-terminal amide
A peptide ending in -CONH₂ rather than -COOH. Frequently required for activity; its hydrolysis to the free acid is a +0.984 Da impurity.
Lactam bridge
A cyclisation through an amide bond between a side-chain carboxyl and a side-chain amine. Melanotan II and bremelanotide are lactam-cyclised.
Macrolactamisation
The cyclisation step forming a lactam bridge. Incomplete cyclisation leaves the linear precursor at +18.011 Da — the commonest process failure in the melanocortin class.
see also Lactam bridge
PEGylation
Covalent attachment of polyethylene glycol. The polymer is polydisperse, so a PEGylated peptide has no single mass and no meaningful single purity figure.
Polydispersity
A distribution of molecular masses rather than a single value. Characteristic of synthetic polymers and therefore of PEG conjugates.
see also PEGylation
Maleimide
A thiol-reactive group used for covalent conjugation, as in CJC-1295 DAC. Hydrolyses to an inert maleamic acid at +18.011 Da — pure and non-functional are compatible states.
Pseudoproline
A dipeptide building block that temporarily disrupts secondary structure on-resin, used to get through aggregation-prone sequences.
Resin
The insoluble polymer support on which solid-phase synthesis is performed. Loading and swelling behaviour determine achievable purity for difficult sequences.
Solid-phase peptide synthesis (SPPS)
Stepwise chain assembly on an insoluble support, with excess reagent driving each coupling. The origin of almost every peptide in this index.
Cleavage cocktail
The acidic mixture that releases the peptide from the resin and removes side-chain protection. Scavengers in it suppress alkylation artefacts.
see also tBu residual
Preparative HPLC
Chromatography at scale to purify the crude peptide. The fraction cut sets the purity/yield trade, and a wide cut is visible in the impurity profile.
Salt exchange
Converting a peptide from one counter-ion to another, typically trifluoroacetate to acetate, by repeated lyophilisation from the new acid or by ion exchange. Rarely complete.
Lyophilisation
Freeze-drying: freezing the solution then subliming the ice under vacuum. Cake structure and residual moisture are the observable outputs.
Collapse temperature
The temperature above which a freeze-concentrated matrix loses structure during primary drying. Exceeding it gives a collapsed cake and high residual moisture.
Melt-back
A lyophilisation failure in which the cake structure is entirely lost, leaving a powder or glassy residue. Expect high residual moisture.
Metal complex
A peptide coordinating a metal ion, as in GHK-Cu and thymulin. An acidic reverse-phase mobile phase dissociates the complex on-column.
see also ICP-MS
Recombinant
Produced by expression in a host organism rather than by chemical synthesis. Brings folding, glycosylation and host-cell-protein questions that a purity figure does not address.
Host cell protein
Protein from the expression organism carried through purification. A required specification for recombinant products and reported on none in this index.
see also Recombinant
14 terms

Physical

Cake
The porous solid remaining after lyophilisation. Its appearance is a direct readout of cycle control.
Cake shrinkage
Separation of the cake from the vial wall. Indicates moisture uptake after drying, or a closure-integrity failure.
see also Cake
Closure integrity
Whether the stopper and crimp maintain the container's seal and vacuum. A failure lets in moisture and oxygen.
Headspace oxygen
Oxygen in the gas space above the product. Correlates with sulfoxide content and is essentially never measured.
Bacteriostatic water
Water containing 0.9 % benzyl alcohol as a preservative. The benzyl alcohol absorbs strongly in the UV and adds peaks to a chromatogram.
Reconstitution
Dissolving a lyophilised cake in diluent. Procedure matters: directing a stream at the cake or shaking can nucleate aggregates that were not in the vial.
Reconstitution clarity
The visual appearance of the reconstituted solution. Recorded verbatim in this index because the descriptive detail is the diagnostic content.
Cold chain
Maintaining a controlled low temperature through shipping and storage. Its failure has a specific analytical signature: oxidation and, for disulfides, reduction.
see also Oxidation
Desiccant
A moisture-absorbing material included in packaging. Its presence or absence changes the water content measured on arrival for hygroscopic material.
Freeze–thaw
Repeated freezing and thawing of a reconstituted solution. A common aggregation trigger and a standard forced-degradation stress.
Hygroscopic
Readily absorbing atmospheric moisture. Proline-rich short peptides and NAD+ are the worst offenders in this index.
Particulate, sub-visible
Particles below the visible threshold, counted by light obscuration under USP <788>. Not measured anywhere in this index.
Photostability
Resistance to light. Tryptophan-containing peptides and the quinolinium adjuncts degrade in clear glass; amber glass is the mitigation.
see also Oxidation
Vial fill volume
The nominal liquid volume filled before lyophilisation. Determines cake geometry and, with concentration, the mass per vial.
see also Fill weight
16 terms

Quality system

Accuracy
Closeness of a measured value to the true value. Demonstrated by recovery experiments during validation.
see also Validation · Precision
Precision
Closeness of repeated measurements to each other. Says nothing about accuracy: a precise method can be consistently wrong.
Certificate of analysis (COA)
A document stating test results for a specific lot. In this market frequently lacking a method section, a lot number, or both.
Chain of custody
The documented sequence of possession of a sample from acquisition to analysis. Determines what a result can be attributed to.
see also Provenance
Linearity
Proportionality of detector response to concentration across the working range. Required for an assay; irrelevant to area percent.
see also Validation · Range
Range
The concentration interval over which accuracy, linearity and precision have been demonstrated.
see also Linearity
Limit of detection (LOD)
The lowest amount detectable, conventionally at a signal-to-noise ratio of 3.
Limit of quantitation (LOQ)
The lowest amount quantifiable with acceptable precision, conventionally S/N ≥ 10 with RSD ≤ 10 %.
Robustness
A method's tolerance to deliberate small changes in conditions — temperature, pH, organic fraction.
see also Validation
Specificity
Demonstration that the response is due to the analyte and nothing else. The hardest characteristic to establish for peptides because of mass-identical isomers.
Specification
A decision about what result is acceptable. Not a measurement. Conflating the two is how "≥ 99.0 %" becomes "99.0 %" on a certificate.
Validation
Documented demonstration that a method is fit for its purpose, per ICH Q2(R2). Not published for any supplier method in this index.
Reference standard
A material of known content used to quantify an analyte. Absent for most research peptides, which is why content is rarely measured.
Retest date
The date by which a material should be re-tested rather than discarded. Distinct from an expiry date and less often stated.
Out of specification (OOS)
A result outside the acceptance criteria. Requires investigation before it can be invalidated; a result cannot be discarded because it is inconvenient.
see also Specification
Pharmacopoeial monograph
An official specification and method for a named substance. Exists for oxytocin, desmopressin, gonadorelin, leuprolide, triptorelin and glucagon among the compounds here.
see also Specification
23 terms

Statistics

Bland–Altman analysis
Paired comparison of two measurement methods by plotting difference against mean. Yields a bias and limits of agreement. Used for inter-lab agreement here.
Bias
A systematic offset between a measurement and a reference. Distinct from scatter: bias in one direction across many observations is a different finding from random disagreement.
Limits of agreement
Mean difference ± 1.96 standard deviations of the difference. The interval within which 95 % of paired differences fall.
Bootstrap
Estimating a confidence interval by resampling the observed data. Useful where a supplier cell is too small for a parametric interval.
Confidence interval
A range that would contain the true value in a stated proportion of repeated samples. Wide intervals on small cells are the honest presentation.
see also n=
Digit preference
Non-uniform distribution of terminal digits in reported figures. A real integrated area does not land on a round number repeatedly.
see also Rule AX-R04
Interquartile range (IQR)
The difference between the 75th and 25th percentiles. The index's preferred spread statistic because purity distributions are left-skewed.
see also Median · Quantile
Median
The 50th percentile. Preferred over the mean throughout this index because purity distributions have long left tails.
Median absolute deviation (MAD)
The median of absolute deviations from the median, scaled by 1.4826 to estimate σ for normal data. Robust to outliers.
see also Median
Quantile
A value below which a stated proportion of observations fall. The index uses the type 7 definition and says so.
Outlier
An observation beyond the Tukey fences at 1.5 IQR from the quartiles. Plotted individually on every box plot here rather than discarded.
see also Tukey fence
Tukey fence
Q1 − 1.5 IQR and Q3 + 1.5 IQR. The convention for defining box-plot whiskers and outliers.
see also Outlier
Percentage point (pp)
The unit of a difference between two percentages. 98.7 % and 97.9 % differ by 0.8 pp, not 0.8 %.
Power
The probability of detecting an effect of a given size if it exists. Small supplier cells here have low power and the index states so rather than concluding no difference.
see also n=
Regression to the mean
The tendency of an extreme observation to be followed by a less extreme one. Why a retest after a bad result usually looks better regardless of any change.
Repeatability
Precision under identical conditions in one session, same analyst, same instrument. The narrowest kind of precision.
Intermediate precision
Precision across days, analysts and instruments within one laboratory. Wider than repeatability and more realistic.
Reproducibility
Precision across laboratories. The index estimates it from paired same-batch comparisons.
Selection bias
Distortion arising because the sample was not drawn at random. The index's largest unmeasured confounder: lots that get tested are lots someone chose to test.
see also Survivorship
Survivorship
A related bias: the material that reaches a laboratory is not representative of material that does not. Compounded here by who chooses to submit and when.
Standard uncertainty (u)
The uncertainty of a measurement expressed as a standard deviation. Combined across sources in quadrature.
Expanded uncertainty (U)
Standard uncertainty multiplied by a coverage factor, usually k = 2 for approximately 95 % confidence.
Uncertainty budget
An itemised account of every contribution to a measurement's uncertainty. No supplier certificate in this index contains one.