Purity by reverse-phase HPLC (% area)
The index reference method RG-2, what it measures, what it does not, and why a purity figure without a stated gradient is uninterpretable.
The injections that establish a chromatographic system was fit to produce the result, and the parameters the index checks before accepting a purity figure.
System suitability is the check that the instrument, column and method were performing acceptably at the moment the sample was injected. Without it, a result describes an unknown system. It is required by every pharmacopoeia and it is documented on almost no research-peptide certificate.
| Parameter | Definition | Index requirement |
|---|---|---|
| Plate count N | N = 5.54 (t/w½)² | ≥ 20,000 for the main peak |
| Asymmetry As | width at 10 % height, ratio of rear to front segment | 0.80 ≤ As ≤ 1.50 |
| Tailing factor T | (a + b) / 2a at 5 % height | T ≤ 2.0 |
| Resolution R | R = 2(t₂ − t₁) / (w₁ + w₂) | ≥ 1.20 to nearest named impurity; ≥ 1.5 preferred |
| Repeatability | RSD of peak area, replicate injections | ≤ 1.0 % (n = 5) |
| Signal-to-noise at LOQ | peak height / peak-to-peak baseline noise | ≥ 10 |
| Retention repeatability | RSD of retention time | ≤ 1.0 % |
Main peak: t = 18.42 min, w½ = 0.212 min
N = 5.54 × (18.42 / 0.212)²
= 5.54 × (86.89)²
= 5.54 × 7 550
= 41 830 ✓ above 20 000
Nearest impurity: t = 17.60 min, w½ = 0.204 min
Convert half-height widths to baseline widths (w ≈ 1.70 × w½ for a Gaussian):
w₁ = 0.361 min, w₂ = 0.347 min
R = 2 × (18.42 − 17.60) / (0.361 + 0.347)
= 2 × 0.82 / 0.708
= 2.32 ✓ well resolved
Same impurity on a 15-minute gradient (retention gap scales down):
gap = 0.82 × 0.5 = 0.41 min, widths roughly halved to 0.181 / 0.174
R = 2 × 0.41 / 0.355 = 2.31
— resolution is approximately preserved for a resolved pair.
The pairs that fail on a short gradient are those separated by
diastereomeric selectivity, which is not preserved.Cited because they are the documents the acceptance criteria above are taken from. The index applies them as written and states every deviation.