AssayIndex

AX-2025-018817

Kisspeptin-10 · 10 mg declared · Qingdao Sigma Chemical Co., Ltd. · lot QSC-2412-W8004

MARGINALPURITY ONLY
Report ID
AX-2025-018817
Compound
Kisspeptin-10
Compound class
Melanocortin-receptor peptides
Supplier
QSC — Qingdao Sigma Chemical Co., Ltd.
Batch / lot
QSC-2412-W8004
Declared label claim
10mg
Laboratory
Nordanalyt Laboratories
Assay date
08 Jan 2025
Submitted
12 Jan 2025
Quarter
2025-Q1
Submitter
so.beaulieu
Sample provenance
Supplier-supplied
Section 1

Results

Measured against the index acceptance specification for the melanocortin-receptor peptides class. A specification is a decision the index made; a result is a measurement. See where each threshold came from.

ParameterMethodResultUnitIndex spec
PurityRP-HPLC, 35 min gradient, 214 nm99.93% area≥ 98.5PASS
Total impuritiesRP-HPLC, sum of reported peaks0.07% area≤ 1.5PASS
Largest single impurityRP-HPLC — des-Tyr (−163 Da)0.04% area≤ 1.00PASS
Identitynot ordered on this submissionnot determinedrequired for VERIFIED statusNOT TESTED
Total peptide contentby difference from the gravimetric mass balance89.17%w/wNOT TESTED
Target-compound contenttotal peptide content × purity89.11%w/wNOT TESTED
Gross fill (as weighed)five-place balance, under nitrogen11.09mg10 mg declaredNOT TESTED
Recovered target peptidegross fill × target content9.882mg10 mg claimNOT TESTED
Recovery against label claimrecovered ÷ claim98.8%90.0 – 110.0PASS
Water contentKarl Fischer, coulometric, duplicate5.89%≤ 8.0PASS
Counter-ion (acetate)anion-exchange ion chromatography4.26%≤ 10.0PASS
Residual solvent + inorganic residueestimated — see /methods/residual-solvents-q3c/0.68%NOT TESTED
Endotoxinnot ordered on this submissionnot determined≤ 5.0NOT TESTED
Sterilitynot ordered on this submissionnot determinedno growthNOT TESTED
Appearancevisual, diffuse light, black and white backgroundWhite lyophilised cake, intactwhite to off-white intact cakePASS
Reconstitution clarityreconstituted per submitter procedureClear with a single visible fibreclear, no visible particulatesFAIL
pH (reconstituted, 1 mg/mL)potentiometric, two-point calibration5.922.5 – 8.0PASS
Retention time (observed)RP-HPLC, 35 min gradient10.56min10.54 ± 0.25 (normalised)PASS
Peak asymmetry (10 % height)Ph. Eur. 2.2.461.220.80 – 1.50PASS
Plate count, main peakN = 5.54 (t/w½)²80,001≥ 20,000PASS
Resolution to nearest impurityvs [D-Trp]-epimer0.91no index criterionREPORTED
Section 2

Chromatogram and peak integration

The figure is drawn from the table. It is not an independent illustration — if a number in the table changed, the trace would change with it.

027915583837411165 Kisspeptin-1010.56 min · 99.93 % 9.91 same trace · ×200 vertical expansion, baseline-corrected · main peak off-scale reporting threshold 0.05 % area 0510152025303540 Retention time (min) Response (mAU)
Figure 1. RP-HPLC trace, 35 min gradient, 214 nm. Drawn from the peak-integration table below: peak positions are the reported retention times, peak areas are the reported area percentages, peak widths follow the reported plate count of 80,001, and skew follows the reported asymmetry factor of 1.22. Baseline noise is seeded from the report ID and is illustrative. The lower panel is the same trace at a stated vertical expansion with the slow gradient drift subtracted, because at full scale a 0.04 % impurity against a 99.93 % main peak is a line two pixels tall; the main peak runs off-scale there by design.
#PeakRTArea (µV·s)% areaΔ mass (Da)
1Kisspeptin-10 — main peak10.564,856,90899.93
2des-Tyr (−163 Da)9.911,9440.04-163.063
3[D-Trp]-epimer10.421,4580.030.000
Closure check — purity + Σ impurities 100.00CLOSES

Reporting threshold 0.05 % area. Peaks below it are not reported and are therefore excluded from the total, which overstates purity by their sum — see peak integration.

Section 4

Gravimetric mass balance

Water, counter-ion, residual and total peptide content are exhaustive and mutually exclusive categories. They must account for 100 % of the cake mass. The index closes this to within 0.02 pp and the build asserts it.

ComponentDetermination%w/w of cake
WaterKarl Fischer5.89
Counter-ion (acetate)ion chromatography4.26
Residual solvent + inorganic residueestimated0.68
Total peptide contentby difference89.17
Total100.00 CLOSES

Supplier certificate, as declared

The supplier's own figures for this lot, beside the independent measurement. Divergence is assessed against the batch median, not against this single report — see rule AX-R05.

FigureSupplier COAThis reportIndex note
Purity (% area)99.9099.93divergence +0.33 pp vs batch median
Total impurities (% area)not stated0.07
Largest single impurity (% area)not stated0.04
Water (%)not stated5.89
Counter-ion (%)not stated4.26
Peptide content (%w/w)91.789.17
Endotoxin (EU/mg)0.32not determined
Method stated on certificateYEScolumn, gradient and wavelength given
Section 5

Provenance and caveats

Where a sample came from changes the weight of the evidence. The index treats it as a first-class field.

Sample provenance
Supplier-supplied
Evidentiary weight
Weaker evidence. Read with the caveat below.
Chain of custody
Supplier despatch → laboratory direct (courier, cold pack, 10 d) → assay. Submitter did not handle the vial.
Assay date
08 Jan 2025
Submitted to the index
12 Jan 2025
Laboratory
Nordanalyt Laboratories, Uppsala, Sweden
Method basis
35 min gradient · high-resolution MS
Verification status
PURITY ONLY
Section 6

Annotations

No annotations on this record. Annotations are short signed technical comments from index contributors; see how to contribute.

Section 7

Related records

Same batch — lot QSC-2412-W8004 n=9

ReportLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2025-018810QSC-2412-W8004Janoshik2025-Q199.765.2190.7Supplier-supplied1MARGINAL
AX-2025-018811QSC-2412-W8004VendorInvestigate2025-Q199.095.35101.5Buyer-sourced1MARGINAL
AX-2025-018812QSC-2412-W8004PeptideMeter2025-Q199.855.96Supplier-supplied0PASS
AX-2025-018813QSC-2412-W8004PeptideMeter2025-Q199.295.21Supplier-supplied1MARGINAL
AX-2025-018814QSC-2412-W8004Janoshik2025-Q199.575.4797.9Supplier-supplied1MARGINAL
AX-2025-018815QSC-2412-W8004PeptideMeter2025-Q199.815.70Supplier-supplied1MARGINAL
AX-2025-018816QSC-2412-W8004Nordanalyt2025-Q199.535.4892.7Supplier-supplied1MARGINAL
AX-2025-018818QSC-2412-W8004Nordanalyt2025-Q199.576.1698.8Supplier-supplied1MARGINAL
Other recent Kisspeptin-10 records in the index.
ReportSupplierLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-042461SWBSWB-2603-D6238Janoshik2026-Q398.354.6596.5Supplier-supplied3FAIL
AX-2026-043176GL BiochemGLBI-2608-N8085Medutest2026-Q398.223.3092.5Buyer-sourced2FAIL
AX-2026-043180GL BiochemGLBI-2608-N8085Janoshik2026-Q398.693.37103.8Supplier-supplied2MARGINAL
AX-2026-042460SWBSWB-2603-D6238Kestrel2026-Q399.554.2190.8Supplier-supplied3MARGINAL
AX-2026-042458SWBSWB-2603-D6238Kestrel2026-Q398.514.71102.4Supplier-supplied3MARGINAL
AX-2026-043178GL BiochemGLBI-2608-N8085Janoshik2026-Q398.244.4496.0Supplier-supplied2FAIL
AX-2026-042463SWBSWB-2603-D6238Nordanalyt2026-Q398.354.9398.6Supplier-supplied3FAIL
AX-2026-042465SWBSWB-2603-D6238Kestrel2026-Q399.034.8396.4Supplier-supplied3MARGINAL
Other recent QSC records in the index.
ReportCompoundLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-041683CJC-1295 (no DAC)QSC-2606-V8637Janoshik2026-Q399.933.6397.1Supplier-supplied1FAIL
AX-2026-041727CJC-1295 with DACQSC-2602-E9474Janoshik2026-Q398.886.62105.3Supplier-supplied1MARGINAL
AX-2026-041728CJC-1295 with DACQSC-2602-E9474VendorInvestigate2026-Q398.427.04Buyer-sourced1MARGINAL
AX-2026-041734CJC-1295 with DACQSC-2602-E9474Medutest2026-Q399.437.17105.5Supplier-supplied2FAIL
AX-2026-041652LiraglutideQSC-2605-W8602Nordanalyt2026-Q399.336.55Supplier-supplied0PASS
AX-2026-041730CJC-1295 with DACQSC-2602-E9474PeptideMeter2026-Q398.796.9395.4Supplier-supplied1MARGINAL
AX-2026-041686CJC-1295 (no DAC)QSC-2606-V8637Janoshik2026-Q399.934.04Supplier-supplied0PASS
AX-2026-041653LiraglutideQSC-2605-W8602VendorInvestigate2026-Q398.825.95Buyer-sourced0PASS