Cross-section

Semax by laboratory

The same compound measured by different laboratories on different gradients. Where the medians differ, the question is whether the material differed or the method did — and for Semax, with a class minimum gradient of 15 min, the method is a live explanation.

160 recordsmedian 99.17 %spec 98.5 %
LaboratoryReportsGradientMS resolutionMedian purity %Published bias (pp)Bias-adjustedMethod suitability
Janoshik6730minHigh99.34+0.0099.34ADEQUATE
PeptideMeter3128minHigh99.25-0.0599.30ADEQUATE
Medutest2025minUnit99.09+0.0899.01ADEQUATE
VendorInvestigate1732minHigh98.85-0.1198.96ADEQUATE
Nordanalyt1535minHigh99.17+0.0299.15ADEQUATE
Kestrel1015minUnit98.91+0.2498.67ADEQUATE
97.698.098.598.999.399.7100.1JanoshikJanoshik: Q1 98.96, median 99.34, Q3 99.75, n=67n=67PeptideMeterPeptideMeter: Q1 98.78, median 99.25, Q3 99.59, n=31n=31MedutestMedutest: Q1 98.44, median 99.09, Q3 99.25, n=20n=20VendorInvestigateVendorInvestigate: Q1 98.38, median 98.85, Q3 99.05, n=17n=17NordanalytNordanalyt: Q1 98.73, median 99.17, Q3 99.70, n=15n=15KestrelKestrel: Q1 98.57, median 98.91, Q3 99.29, n=10n=10 Purity (% area)
Figure. Semax purity by contributing laboratory. Any offset between rows has two possible sources — a real difference in the material each laboratory happened to receive, and the laboratory's own bias — and this figure cannot separate them. The paired analysis that can is on inter-lab agreement.