AssayIndex
Cross-section

Ipamorelin by laboratory

The same compound measured by different laboratories on different gradients. Where the medians differ, the question is whether the material differed or the method did — and for Ipamorelin, with a class minimum gradient of 15 min, the method is a live explanation.

223 recordsmedian 99.48 %spec 99.0 %
LaboratoryReportsGradientMS resolutionMedian purity %Published bias (pp)Bias-adjustedMethod suitability
Janoshik10830minHigh99.54+0.0099.54ADEQUATE
PeptideMeter3328minHigh99.28-0.0599.33ADEQUATE
Medutest2725minUnit99.72+0.0899.64ADEQUATE
Kestrel1915minUnit99.84+0.2499.60ADEQUATE
VendorInvestigate1832minHigh99.13-0.1199.24ADEQUATE
Nordanalyt1835minHigh99.41+0.0299.39ADEQUATE
97.197.698.198.699.199.6100.1JanoshikJanoshik: Q1 98.95, median 99.54, Q3 99.93, n=108outlier 97.34n=108PeptideMeterPeptideMeter: Q1 98.74, median 99.28, Q3 99.70, n=33n=33MedutestMedutest: Q1 99.06, median 99.72, Q3 99.93, n=27n=27VendorInvestigateVendorInvestigate: Q1 98.68, median 99.13, Q3 99.44, n=18n=18NordanalytNordanalyt: Q1 99.27, median 99.41, Q3 99.93, n=18outlier 97.94n=18KestrelKestrel: Q1 99.25, median 99.84, Q3 99.93, n=19n=19 Purity (% area)
Figure. Ipamorelin purity by contributing laboratory. Any offset between rows has two possible sources — a real difference in the material each laboratory happened to receive, and the laboratory's own bias — and this figure cannot separate them. The paired analysis that can is on inter-lab agreement.