AssayIndex

AX-2024-005071

TB-500 · 5 mg declared · Xi'an Mokemei Biotechnology Co., Ltd. · lot MKM-2402-N3279

FAILVERIFIED
Report ID
AX-2024-005071
Compound
TB-500
Compound class
Tissue-repair and immunomodulatory peptides
Supplier
MKM — Xi'an Mokemei Biotechnology Co., Ltd.
Batch / lot
MKM-2402-N3279
Declared label claim
5mg
Laboratory
Janoshik Analytical
Assay date
25 Jun 2024
Submitted
03 Jul 2024
Quarter
2024-Q2
Submitter
s.erdogan
Sample provenance
Supplier-supplied
Section 1

Results

Measured against the index acceptance specification for the tissue-repair and immunomodulatory peptides class. A specification is a decision the index made; a result is a measurement. See where each threshold came from.

ParameterMethodResultUnitIndex spec
PurityRP-HPLC, 30 min gradient, 214 nm99.93% area≥ 98.5PASS
Total impuritiesRP-HPLC, sum of reported peaks0.07% area≤ 1.5PASS
Largest single impurityRP-HPLC — [D-Thr]-diastereomer0.04% area≤ 1.00PASS
Identity — theoretical monoisotopic masscalculated from molecular formula887.4964DaNOT TESTED
Identity — observed monoisotopic massLC-MS, high resolution (ESI-TOF/QTOF)887.4950DaNOT TESTED
Mass accuracyobserved vs theoretical-1.6ppm± 15PASS
Peptide contentnot ordered on this submissionnot determinedrequired to check label claimNOT TESTED
Water contentKarl Fischer, coulometric, duplicate3.91%≤ 10.0PASS
Counter-ion (acetate)anion-exchange ion chromatography6.05%≤ 12.0PASS
Residual solvent + inorganic residueestimated — see /methods/residual-solvents-q3c/0.32%NOT TESTED
Endotoxinkinetic chromogenic LAL11.28EU/mg≤ 5.0FAIL
Sterilitynot ordered on this submissionnot determinedno growthNOT TESTED
Appearancevisual, diffuse light, black and white backgroundOff-white lyophilised cake, intactwhite to off-white intact cakePASS
Reconstitution clarityreconstituted per submitter procedureFaint haze, clearing on standingclear, no visible particulatesFAIL
pH (reconstituted, 1 mg/mL)potentiometric, two-point calibration5.772.5 – 8.0PASS
Retention time (observed)RP-HPLC, 30 min gradient7.84min7.86 ± 0.25 (normalised)PASS
Peak asymmetry (10 % height)Ph. Eur. 2.2.461.170.80 – 1.50PASS
Plate count, main peakN = 5.54 (t/w½)²64,989≥ 20,000PASS
Resolution to nearest impurityvs [D-Thr]-diastereomer0.87no index criterionREPORTED
Section 2

Chromatogram and peak integration

The figure is drawn from the table. It is not an independent illustration — if a number in the table changed, the trace would change with it.

0346769331040013867 TB-5007.84 min · 99.93 % 7.73 same trace · ×200 vertical expansion, baseline-corrected · main peak off-scale reporting threshold 0.05 % area 05101520253035 Retention time (min) Response (mAU)
Figure 1. RP-HPLC trace, 30 min gradient, 214 nm. Drawn from the peak-integration table below: peak positions are the reported retention times, peak areas are the reported area percentages, peak widths follow the reported plate count of 64,989, and skew follows the reported asymmetry factor of 1.17. Baseline noise is seeded from the report ID and is illustrative. The lower panel is the same trace at a stated vertical expansion with the slow gradient drift subtracted, because at full scale a 0.04 % impurity against a 99.93 % main peak is a line two pixels tall; the main peak runs off-scale there by design.
#PeakRTArea (µV·s)% areaΔ mass (Da)
1TB-500 — main peak7.844,856,90899.93
2[D-Thr]-diastereomer7.731,9440.040.000
3C-terminal amide (−1 Da)8.021,4580.03-0.984
Closure check — purity + Σ impurities 100.00CLOSES

Reporting threshold 0.05 % area. Peaks below it are not reported and are therefore excluded from the total, which overstates purity by their sum — see peak integration.

Section 3

Mass spectrum and identity

Theoretical monoisotopic mass 887.4964 Da, computed from C39H69N9O14. All ppm figures on this site are against monoisotopic values — see why that matters.

0255075100375475575675775875975 [M+1H]1+888.50[M+2H]2+444.75 m/z Relative abundance (%)
Figure 2. ESI charge-state envelope, high-resolution instrument. Sticks are placed at m/z = (M + z × 1.007276)/z for the observed neutral mass of 887.4950 Da. Sodium (+21.982/z) and potassium (+37.956/z) adducts are shown where the parent charge state exceeds 25 % relative abundance; they are adducts of the same molecule and are not impurities.
ChargeTheoretical m/zObserved m/zΔ m/zRel. abundance (%)
2+444.7555444.7548-0.000750.7
1+888.5037888.5023-0.0014100.0
Section 4

Gravimetric mass balance

Water, counter-ion, residual and total peptide content are exhaustive and mutually exclusive categories. They must account for 100 % of the cake mass. The index closes this to within 0.02 pp and the build asserts it.

Supplier certificate, as declared

The supplier's own figures for this lot, beside the independent measurement. Divergence is assessed against the batch median, not against this single report — see rule AX-R05.

FigureSupplier COAThis reportIndex note
Purity (% area)99.9099.93divergence +0.24 pp vs batch median
Total impurities (% area)0.100.07
Largest single impurity (% area)not stated0.04
Water (%)not stated3.91
Counter-ion (%)not stated6.05
Peptide content (%w/w)not statednot determined
Endotoxin (EU/mg)not stated11.28
Method stated on certificateNOno method section — the result cannot be reconciled against any other
Section 5

Provenance and caveats

Where a sample came from changes the weight of the evidence. The index treats it as a first-class field.

Sample provenance
Supplier-supplied
Evidentiary weight
Weaker evidence. Read with the caveat below.
Chain of custody
Supplier despatch → laboratory direct (courier, cold pack, 11 d) → assay. Submitter did not handle the vial.
Assay date
25 Jun 2024
Submitted to the index
03 Jul 2024
Laboratory
Janoshik Analytical, Ostrava, Czechia
Method basis
30 min gradient · high-resolution MS
Verification status
VERIFIED
Section 6

Annotations 4

Signed technical comments from index contributors. Not a discussion forum — annotations question or contextualise the result, and the data team replies where a reply is warranted.

pegmgf_disperse
Endotoxin determined but the LAL method is not stated. Kinetic chromogenic and gel-clot give different results and different sensitivity. The value cannot be interpreted without knowing which.
follistatin_gap
11.28 EU/mg from an older lot. Endotoxins are heat-stable and drift is rare. If this matches fresh-lot median from the same supplier, the number is stable.
valley_drop
Endotoxin assay but no bacterial viability or total count. The two are independent: dead endotoxin-bearing bacteria contribute to the burden, living ones that produce endotoxins do not.
cjc_maleimide
11.28 EU/mg is at the sensitivity limit of Janoshik's LAL method. The true value is unspecified; treating this as a point estimate is overstating precision.
Section 7

Related records

Same batch — lot MKM-2402-N3279 n=8

ReportLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2024-005066MKM-2402-N3279PeptideMeter2024-Q299.724.85Supplier-supplied2FAIL
AX-2024-005067MKM-2402-N3279VendorInvestigate2024-Q299.374.68Buyer-sourced1MARGINAL
AX-2024-005068MKM-2402-N3279Medutest2024-Q299.723.94Supplier-supplied1MARGINAL
AX-2024-005069MKM-2402-N3279Janoshik2024-Q299.924.52Supplier-supplied1MARGINAL
AX-2024-005070MKM-2402-N3279Janoshik2024-Q299.594.49101.9Supplier-supplied1MARGINAL
AX-2024-005072MKM-2402-N3279VendorInvestigate2024-Q299.434.59Laboratory-retained1MARGINAL
AX-2024-005073MKM-2402-N3279Nordanalyt2024-Q299.394.0990.3Supplier-supplied1MARGINAL
Other recent TB-500 records in the index.
ReportSupplierLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-041847JEEPJEEP-2602-V3434Nordanalyt2026-Q399.723.85Supplier-supplied1MARGINAL
AX-2026-042574ERPERP-2605-S8022VendorInvestigate2026-Q399.542.9196.4Buyer-sourced1MARGINAL
AX-2026-042577ERPERP-2605-S8022Janoshik2026-Q399.513.45Supplier-supplied1MARGINAL
AX-2026-041845JEEPJEEP-2602-V3434Janoshik2026-Q399.724.04102.2Supplier-supplied1MARGINAL
AX-2026-041921JEEPJEEP-2605-M4874Janoshik2026-Q399.901.76Supplier-supplied2MARGINAL
AX-2026-041922JEEPJEEP-2605-M4874Janoshik2026-Q399.412.6095.7Laboratory-retained2MARGINAL
AX-2026-041928JEEPJEEP-2605-M4874Janoshik2026-Q399.881.8492.8Laboratory-retained2MARGINAL
AX-2026-041849JEEPJEEP-2602-V3434VendorInvestigate2026-Q399.733.6088.6Buyer-sourced2MARGINAL
Other recent MKM records in the index.
ReportCompoundLotLabQuarterPurity %Water %Recovery %ProvenanceFlagsResult
AX-2026-042943AdipotideMKM-2606-K4013Janoshik2026-Q397.926.31Supplier-supplied1MARGINAL
AX-2026-042891CJC-1295 (no DAC)MKM-2603-V2347Janoshik2026-Q399.283.63Supplier-supplied1MARGINAL
AX-2026-042942AdipotideMKM-2606-K4013VendorInvestigate2026-Q397.045.5694.6Buyer-sourced1FAIL
AX-2026-042914GonadorelinMKM-2604-A2596Janoshik2026-Q299.935.19Supplier-supplied0PASS
AX-2026-042884BPC-157MKM-2605-X2575Nordanalyt2026-Q298.122.89Laboratory-retained1MARGINAL
AX-2026-042946TirzepatideMKM-2601-L3752Medutest2026-Q297.718.10Supplier-supplied1FAIL
AX-2026-042911GonadorelinMKM-2604-A2596Medutest2026-Q299.085.55Buyer-sourced1MARGINAL
AX-2026-042932L-Glutathione (reduced)MKM-2511-F2902Janoshik2026-Q298.578.4192.9Supplier-supplied1MARGINAL